CVE-2019-25722

HIGH CVSS 4.0: 7.2 EPSS 0.19%
Updated Jul 22, 2026
Dr
Parameter Value
CVSS 7.2 (HIGH)
Type CWE-798 (Hardcoded Credentials)
Vendor Dr
Public PoC No

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

Attack Parameters

Attack Vector
Adjacent
Requires local network access
Attack Complexity
Low
Easy to exploit
Attack Requirements
None
No additional conditions
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
Low
Partial data leak
Integrity
Low
Partial data modification
Availability
High
Complete denial of service