CVE-2019-25723

MEDIUM CVSS 4.0: 6.3 EPSS 0.24%
Updated Jul 22, 2026
Dr
Parameter Value
CVSS 6.3 (MEDIUM)
Affected Versions 2.00 — 2.02
Type CWE-1286
Vendor Dr
Public PoC No

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Attack Parameters

Attack Vector
Network
Can be exploited remotely
Attack Complexity
High
Difficult to exploit
Attack Requirements
None
No additional conditions
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
None
No data leak
Integrity
None
No data modification
Availability
Low
Partial disruption

CVSS Vector v4.0

Weakness Type (CWE)