CVE-2021-4479

MEDIUM CVSS 4.0: 6.3 EPSS 0.24%
Updated Jul 22, 2026
Dr
Parameter Value
CVSS 6.3 (MEDIUM)
Type CWE-1286
Vendor Dr
Public PoC No

Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

Attack Parameters

Attack Vector
Network
Can be exploited remotely
Attack Complexity
High
Difficult to exploit
Attack Requirements
None
No additional conditions
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
None
No data leak
Integrity
None
No data modification
Availability
Low
Partial disruption

CVSS Vector v4.0

Weakness Type (CWE)