CVE-2022-4992

HIGH CVSS 4.0: 8.8 EPSS 0.16%
Updated Jul 22, 2026
Dr
Parameter Value
CVSS 8.8 (HIGH)
Type CWE-345 (Insufficient Verification of Data)
Vendor Dr
Public PoC No

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Attack Parameters

Attack Vector
Network
Can be exploited remotely
Attack Complexity
Low
Easy to exploit
Attack Requirements
None
No additional conditions
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
Low
Partial data leak
Integrity
Low
Partial data modification
Availability
High
Complete denial of service

CVSS Vector v4.0