CVE-2025-0683

HIGH CVSS 4.0: 8.2 EPSS 0.78%
Updated Jan 31, 2025
In
Parameter Value
CVSS 8.2 (HIGH)
Type CWE-359
Vendor In
Public PoC No

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Attack Parameters

Attack Vector
Network
Can be exploited remotely
Attack Complexity
Low
Easy to exploit
Attack Requirements
Present
Additional conditions required
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
High
Complete data leak
Integrity
None
No data modification
Availability
None
No disruption

CVSS Vector v4.0