CVE-2025-14095

MEDIUM CVSS 3.1: 6.8 EPSS 0.18%
Updated Oct 07, 2026
Parameter Value
CVSS 6.8 (MEDIUM)
Type CWE-693, CWE-284 (Improper Access Control)
Public PoC No

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.

Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Attack Parameters

Attack Vector
Physical
Requires physical access
Attack Complexity
Low
Easy to exploit
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
High
Complete data leak
Integrity
High
Complete data modification
Availability
High
Complete denial of service

CVSS Vector v3.1