CVE-2025-15653

HIGH CVSS 4.0: 7.0 EPSS 0.17%
Updated Jul 22, 2026
Dr
Parameter Value
CVSS 7.0 (HIGH)
Type CWE-668 (Exposure of Resource to Wrong Sphere)
Vendor Dr
Public PoC No

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

Attack Parameters

Attack Vector
Physical
Requires physical access
Attack Complexity
Low
Easy to exploit
Attack Requirements
None
No additional conditions
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
High
Complete data leak
Integrity
High
Complete data modification
Availability
High
Complete denial of service

CVSS Vector v4.0