CVE-2026-0611

CRITICAL CVSS 4.0: 9.2 EPSS 0.66%
Updated Jul 22, 2026
Spacelabs Healthcare
Parameter Value
CVSS 9.2 (CRITICAL)
Affected Versions before 11.6.0
Type CWE-306 (Missing Authentication for Critical Function)
Vendor Spacelabs Healthcare
Public PoC No

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.

Attack Parameters

Attack Vector
Network
Can be exploited remotely
Attack Complexity
Low
Easy to exploit
Attack Requirements
Present
Additional conditions required
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
High
Complete data leak
Integrity
High
Complete data modification
Availability
High
Complete denial of service

CVSS Vector v4.0