CVE-2026-7251

CRITICAL CVSS 4.0: 9.3 EPSS 0.50%
Updated Jul 24, 2026
Eppendorf
Parameter Value
CVSS 9.3 (CRITICAL)
Type CWE-259
Vendor Eppendorf
Public PoC No

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320.

VNC traffic is not encrypted.

Attack Parameters

Attack Vector
Network
Can be exploited remotely
Attack Complexity
Low
Easy to exploit
Attack Requirements
None
No additional conditions
Privileges Required
None
No privileges needed
User Interaction
None
No user interaction needed

Impact Assessment

Confidentiality
High
Complete data leak
Integrity
High
Complete data modification
Availability
High
Complete denial of service

CVSS Vector v4.0

Weakness Type (CWE)